LYRA x1 Most performing and versatile ventilation for hospital applications
LYRA x1 Most performing and versatile ventilation for hospital applications
Vizualizari:
1781
Cod:
683
Producator:
Axcent Germania
Cod producator:
LYRA x1
In stoc:
Nu
Garantie:
24 luni
0.00
Lei
Comanda rapida
Nume:
*
E-mail:
*
*
Telefon:
*
Cantitate:
*
Detalii produs
TEHNICAL SPECIFICATIONS
Physical Specification
Dimensions: 336 mm x 330 mm x 345 mm (L x W x H): 664 mm x 600 mm x 1370 mm (with trolley)
Weight: Approximately 9.5 kg,
Approximately 31.0 kg (with trolley)
Screen
Display Size: 12.1 Color active matrix TFT touch
Display Resolution (H) x (V): 1280 x 800 pixels
Brightness: Adjustable
Ventilation Specifications
Patient Type: Adult, Pediatic
Invasive Ventilation Mode:
VCV
(Volume Control Ventilation)
PCV
(Pressure Control Ventilation)
VSIMV
(Volume Synchronized Intermittent Mandatory Ventilation)
PSIMV
(Pressure Synchronized Intermittent Mandatory Ventilation)
CPAP/PSV
(Continuous Positive Airway Pressure/Pressure Support Ventilation)
PRVC
(Pressure Regulated Volume Control)
V + SIMV
(PRVC + SIMV)
Apnea Ventilation
Non-invasive Ventilation Mode:
PCV
(Pressure Control Ventilation)
PSIMV
(Pressure Synchronized Intermittent Mandatory Ventilation )
CPAP/PSV
(Continuous Positive Airway
Pressure/Pressure Support Ventilation)
BPAP
(Bilevel Positive Airway Pressure)
APRV
(Airway Pressure Release Ventilation
Controlled Parameters
O
2
%: 21-100% (increments of 1%)
VT (Tidal Volume): Adult: 100-2000 mL (increments of 10 mL) / Pediatric: 20-300 mL (increments of 1 mL)
f (Ventilation frequency): 1-80 bpm (increments of 1 bpm)
fSIMV (Ventilation frequency in SIMV mode): 1-80 bpm (increments of 1 bpm)
I:E range: 4:1-1:10 (increments of 0.5)
Tinsp (Inspiratory time): 0.20-10 s (increments of 0.05 s)
Tslope (Time of Pressure Rising): 0-2.00 s (increments of 0.05 s)
Thigh: 0.2-30 s (increments of 0.1 s)
Tlow: 0.2-30 s (increments of 0.1 s)
Tpause: 5%-60% (increments of 1%), Off)
ΔPinsp: 0-60 cm H
2
O (increments of 1 cm H
2
O)
ΔPsupp: 0-60 cm H
2
O (increments of 1 cm H
2
O)
Phigh: 0-60 cm H
2
O (increments of 1 cm H
2
O)
Plow: 0-45 cm H
2
O (increments of 1 cm H
2
O)
PEEP: 1-45 cm H
2
O (increments of 1 cm H
2
O), Off)
Flow trigger: 0.5-15 L/min (increments of 0.1 L/min)
Pressure trigger: -10 to -0.5 cm H
2
O (increments of 0.5 cm H
2
O)
Exp% (Expiration termination level): 10-85% (increments of 5%), Auto
Apnea Ventilation
Vtapnea: Adult: 100-2000 mL (increments of 10 mL) / Pediatric: 20-300 mL (increments of 1 mL)
ΔPapnea: 5-60 cm H
2
O (increments of 1 cm H
2
O)
Fapnea: 1-80 bpm (increments of 1 bpm)
Apnea Tinsp: 0.20-10 s (increments of 0.05 s)
Sigh
Sigh Switch: On, Off)
Interval: 20 s-180 min (increments of 1 s from 20 to 59 s, increments of 1 min from 1 to 180 min)
Cycles Sigh: 1-20 (increments of 1 )
Δint.PEEP: 1-45 cm H
2
O (increments of 1 cm H
2
O), Off)
Synchronized Tube Resistance Compliance
Tube Type: ET Tube, Trach Tube, Disable STRC
Tube I.D.: Adult: 5.0 -12.0 mm (increments of 0.5 mm) / Pediatric: 2.5 - 8.0 mm (increments of 0.5 mm)
Compensate: 0-100% (increments of 1%)
Expiration Compliance Switch: On, Off).
Monitored parameters
Numeric:
Paw
Ppeak
Pplat
Pmean
PEEP
Insp Flow
Exp Flow
MV
MV leak
MV spn
Vte
VTi
Oxygen concentration
VTe spn
VTe/IBW
ftotal
fmand
fspn
Re Continuous Flow (O Therapy)
Ri
Cdyn
Cstat
Rcexp
WOB
RSBI
NIF
P0.1
PEEPi
Real time Graphics:
Pressure-time waveforms: Paw-Volume Loop
Flow-time waveforms: Flow-time Loop
Volume-time waveforms: Paw-Flow Loop
Control Accuracy
O
2
%: ±(3 vol.% +1% of setting)
TV: ±(10 mL +10% of setting) (BTPS)
Tinsp: ±0.1 s or ±10% of setting, whichever is greater
I: E 2:1 to 1:4: ±10% of setting, other range: ±15% of setting
f: ±1 bpm
fSIMV: ±1 bpm
Tslope: ±(0.2 s + 20% of setting)
PEEP: ±(2.0 cm H
2
O + 5% of setting)
ΔPinsp: ±(2.0 cm H
2
O + 5% of setting)
ΔPsupp: ±(2.0 cm H
2
O + 5% of setting)
Phigh: ±(2.0 cm H
2
O + 5% of setting)
Plow: ±(2.0 cm H
2
O + 5% of setting)
Thigh: ±0.2 s or ±10% of setting, whichever is greater
Tlow: ±0.2 s or ±10% of setting, whichever is greater
Pressure Trigger: ±(1.0 cm H
2
O + 10% of setting)
Flow Trigger: ±(1.0 L/min + 10% of setting)
Δint.PEEP: ±(2.0 cm H
2
O + 5% of setting)
Exp%: ±10%
fapnea: ±1 bpm
ΔPapnea: ±(2.0 cm H
2
O + 5% of setting)
Tvapnea: ±(10 mL + 10% of setting) (BTPS)
Apnea Tinsp: ±0.1 s or ±10% of setting, whichever is greater
Monitoring Accuracy
Airway pressure (Ppeak, Pplat, Pmean, PEEP, PAP, EPAP): ±(2 cm H
2
O + 4% of the actual reading)
Tidal Volume: (Tvi, Tve, TVe/IBW, TVe spn): 0 ml-100 ml: ±(10 ml + 3% of the actual reading) (BTPS) 100 ml-4000 ml: ±(3 ml +10% of the actual reading) (BTPS)
Minute Volume (MV, MVspn, Mvleak): ±0.3 L/min or ±8% of the actual reading, whichever is greater (BTPS)
Frequency (ftotal, fmand, fspn): ±5% of reading or ±1bpm, whichever is greater
Inspired Oxygen (FiO
2
): ±(2.5 vol.% + 2.5% of the actual reading)
Resistance: 0 to 50: ±10 cm H
2
O/L/s
Other range: 50% of the actual reading
Compliance: 25% of the actual reading or ±10 ml/cm H
2
O, whichever is greater
RSBI: 0 to 1000 1/(Lmin): 15% of the actual reading or ±20 1/(Lmin)
WOB: -
NIF: ±(2 cm H
2
O + 4% of the actual reading)
P0.1: ±(2 cm H
2
O + 4% of the actual reading)
PEEPi: -
Rcexp: -
Alarm settings
Tidal Volume: High: Adult: 110-4000 mL, Off / Pediatric: 25-600 mL, Off / Low: Adult: 50-4000 ml, Off / Pediatric: 10-600 mL, Off)
Minute Volume: High: Adult: 0.2-100.0 L/min / Pediatric: 0.2-60.0 L/min / Low: Adult: 0.1-50.0 L/min / Pediatric: 0.1-30.0 L/min
Airway pressure Frequency High: 10-85 cm H
2
O
Inspired oxygen (FiO
2
) High: Actual reading + 8% (High-pressure oxygen), 2%-100%
(Low-pressure oxygen) / Low: Actual reading - 8% (High-pressure oxygen), 18%-98%
(Low-pressure oxygen)
Apnea alarm time: 5-60 s
Trend
Type: Tabular, Graphic
Length: 72 hours
Content: Monitor Parameters, Setting Parameters (Setting Ventilation mode and Parameters
O
2
Therapy
Controlled Parameters
O
2
%: 21-100% (increments of 1%)
Flow: 4-60 L/min
Controlled Accuracy
O
2
%: ±(3 vol.% +1% of setting)
Flow: ±(2 L/min +10% of setting) (BTPS)
Environmental specifications
Temperature: 5-40 °C (operating); -20 to 60 °C (storage and transport, O
2
sensor: -20 to 50 °C )
Relative Humidity: 10-95% (operating); 10-95% (storage and transport)
Barometric Pressure: 62-106 kPa (operating); 50-106 kPa (storage and transport)
Power Battery Backup
External AC power supply
Input voltage: 100-240 V
Input frequency: 50/60 Hz
Input current: 2.5 A Max
Fuse: T2.5 AH/250 V
Internal battery
Number of batteries: One or Two (Optional)
Battery type: Build-in Lithium-ion battery, 11.25 VDC, 6400 mAh
Battery run time: 3 hours (Powered by one new fully-charged battery in standard working condition), 6 hours (powered by two new fully-charged batteries in standard working condition)
Others
Communication interface: Rs232, Ethernet, USB port, Nurse call, CO
2
calibration connector
Gas supply: O
2
Pipe Connector: NIST or DISS
Gas supply pressure: 280-600 kPa
Trolley: Dimensions 1000 mm x 676 mm x 505 mm,
Approximately 21 kg
Va rugam sa ne contactati pentru oferta de pret!
Echipa SANTIMPEX