CETUS xl Advanced Patient Monitor
CETUS xl Advanced Patient Monitor
Vizualizari:
1504
Cod:
682
Producator:
Axcent Germania
Cod producator:
CETUS xl Advanced Patient Monitor
In stoc:
Nu
Garantie:
24 luni
0.00
Lei
Comanda rapida
Nume:
*
E-mail:
*
*
Telefon:
*
Cantitate:
*
Detalii produs
Features
•
15.6”/17/19” switchable TFT LCD Touch Screen
•
Aluminium material shell
•
Fanless design suitable for quite care environment
•
10 waveform display, up to 12-lead ECG analysis
•
Powerful calculation (Hemodynamic, Drug Dose, Oxygenation, Ventilation)
•
SpO2 support PVI and PI, low perfusion 0.2%
•
Aspect module, NMT (Organon TOF-Watch
®
) module optional
•
Wired/Wireless CMS, support HL7 protocol to HIS
•
SpO2 pulse-tone modulation (Pitch Tone)
•
VGA support external display
•
Graphical & tabular trend review (120 hours)
•
48 h full disclosure wave review for each patient (stored in SD card)
Configuration
5-lead ECG, SpO2, NIBP, Resp, PR; Li-ion battery
Optional
12-lead ECG, Masimo/Nellcor SpO2, IBP, C.O., EtCO2, Multi-gas, BIS, NMT; VGA, Thermal Recorder, Wired/Wireless CMS
TEHNICAL SPECIFICATIONS
Display
15.6” TFT Touch screen
Resolution:
1366 x 768
Number of traces:
12 waveforms
I/O
LAN:
1 standard RJ45 port
WLAN:
IEEE 802.11b/g/n
USB:
2 USB connectors
SD:
1 SD card socket
VGA:
1 VGA monitor connector
Output:
1 connector for Nurse call, Defib Sync Analog Output
ECG
Lead type:
3-lead, 5-lead,12-lead
ECG waveform:
2 channels, 7 channels, 12 channels
Display sensitivity:
2.5 mm/mV (×0.25), 5 mm/mV (×0.5), 10 mm/mV (×1.0), 20 mm/mV (×2.0)
Wave sweep speed:
6.25mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s
Bandwidth
Diagnostic mode: 0.05 Hz~100 Hz
Monitor mode: 0.5 Hz~40 Hz
Surgery mode: 1 Hz~20 Hz
Strong filter mode: 5Hz~20 Hz
CMRR>100dB
Notch:
50/60Hz notch filter can be set to on or off
Differential input impedance >5 MΩ
Electrode polarization voltage range: ±400 mV
Baseline recovery time <3s after defibrillation (in monitor and surgery mode)
Calibration signal:
1 mV (peak - peak), accuracy ±3%
RESP
Measurement method:
Thoracic electrical bioimpedance
Measuring lead:
Lead I, II
Wave gain:
×0.25, ×0.5, ×1, ×2
Respiratory impedance range:
0.5-5 Ω
Baseline impedance:
500-4000 Ω
Gain:
10 grades
Scan speed:
6.25 mm/s, 12.5 mm/s, 25 mm/s
TEMP
Accuracy:
±0.1 °C or ±0.2 °F (without probe)
Measurement range:
5~50 °C (41~122 °F)
Channel:
Two channels
Resolution:
0.1 °C
Parameters:
T1, T2 and TD
SpO2
Measurement range:
0-100%
Resolution:
1%
Accuracy:
±2% (70-100%, Adult/Pediatric); ±3% (70-100%, Neonate); 0-69%, unspecified
Refreshing Rate:
1s
Masimo SET® SpO2(Optional)
Measurement range:
0-100%
Resolution:
1%
Accuracy: ±2% (70-100%, Adult/Pediatric, non-motion, low prefusion); ±3% (70-100%, Neonate, non-motion); ±3% (70-100%, motion); 0-69%, unspecified
Refreshing Rate:
1s
Pulse Rate
Range:
30~300 bpm
Resolution:
1bpm
Accuracy:
±2bpm (non-motion), ±5bpm (motion)
Refreshing rate:
1s
NIBP
Measurement method:
Automatic oscillometric method
Operating mode:
Manual, automatic, continuous
Measurement unit:
mmHg/kPa selectable
Typical measurement time:
20~40s
Measurement type:
Systolic, Diastolic, Mean
Measurement range (mmHg)
Range of Systolic pressure:
Adult
40-270
Pediatric 40-200
Neonatal 40-135
Range of Diastolic pressure:
Adult
10-210
Pediatric 10-150
Neonatal 10-95
Range of Mean pressure: Adult 20-230
Pediatric 20-165
Neonatal 20-105
Measurement accuracy
Maximum average error:
±5 mmHg
Maximum standard deviation:
8 mmHg
Resolution:
1 mmHg
Interval:
1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180, 240, 480 minutes
Overpressure protection:
Software and hardware, double safety protection
Cuff pressure range:
0-280mmHg
IBP (Optional)
Channel:
2-channel or 4-channel
ART:
0 to 300 mmHg
PA:
-6 to 120 mmHg
CVP/RAP/LAP/ICP:
-10 to 40 mmHg
Measurement range:
P1/P2 -50 to 300 mmHg
Resolution:
1 mmHg
Accuracy:
±2% or ±1 mmHg, whichever is greater (without sensor)
Sensitivity:
5uV/mmHg/V
Impedance range:
300 to 3000 Ω
C.O. (Optional)
Method:
Thermodilution
Range:
C.O.: 0.2 to 20 L/min, TB: 23 to 45 °C, T1: -1 to 27 °C
Accuracy:
C.O.: ±5% or ±0.1L/min, whichever is greater TB, T1: ±0.5°C (without sensor)
Standard Mainstream CO2 (Optional)
Measurement range:
0-19.7%, 150 mmHg, or 0-20kPa
Resolution:
0.1 mmHg
Measurement accuracy
0 - 40 mmHg: ± 2 mmHg
41 - 70 mmHg: ± 5% of reading
71 - 100 mmHg: ± 8% of reading
101 - 150 mmHg: ± 10% of reading
Respiration rate:
3-150 bpm
Respiration rate accuracy:
1% ±1 bpm
Warm-up time:
97% within 8s, full accuracy within 20s
Standard Sidestream CO2 (Optional)
Measurement rage:
0-20% (0-150 mmHg)
Accuracy:
< 5.0% CO 2: ± 2 mmHg
> 5.0% CO 2: < 6% of reading
Respiration rate:
2~150 BPM
Respiration rate accuracy:
1% ±1BPM
Warm-up time:
97% within 45s, full accuracy within 10 min
Rise times (t10-90%):
About 100 ms, when flow is 100 ml/min, adult water trap?1.5m sampling tube
Delay time:
<3sec when flow is 100 ml/min, adult water trap?1.5 m sampling tube
Recorder (Optional)
Built-in, Thermal dot array
Horizontal resolution:
16 dots/mm (25 mm/s paper speed)
Vertical resolution:
8 dots/mm
Paper speed:
25 mm/s, 50 mm/s
Number of waveform channels:
3
Phasein ISA Sidestream CO2 (Optional)
Warm-up time:
Full accuracy within 10 seconds
Sampling flow rate:
50ml/min(+/-10/min)
Measurement Range:
0 -25%
Accuracy:
0~15% (±0.2% of the reading)
15~25%, unspecified
Rise time:
200 ms, typical at 50 ml/min flow rate
Total response time: within 3 seconds (with 2 m Momoline sampling line)
AWRR Range:
0-150 bpm
AWRR Accuracy:
±1 breath
Phasein IRMA™ Mainstream CO2 (Optional)
Measurement Range:
0 -25%
Accuracy:
0~15% (±0.2% of the reading)
15~25%, unspecified
Warm-up time:
Full accuracy within 10 seconds
AWRR Range:
0-150 bpm
AWRR Accuracy:
±1 breath
Phasein IRMA™ AX+ Mainstream Multi-gas (Optional)
Gas:
CO2, N2O, HAL, ISO, ENF, SEV, DES with automatic identification
Warm-up time:
Full accuracy within 20 seconds for IRMA AX+
CO2 Accuracy:
0-10%: ± (0.2%+2% of the reading)
0-15%: ± (0.3%+2% of the reading)
N2O Accuracy:
0-100%: ± (2%+2% of the reading)
HAL, ISO, ENF:
0-8%: ± (0.15%+5% of the reading)
SEV:
0-10%: ± (0.15%+5% of the reading)
DES:
0-22%: ± (0.15%+5% of the reading)
Agent identification time:
<20s(typical <10s)
AWRR range:
0-150 bpm
AWRR accuracy:
+/-1 bpm
Apnea time:
20~60s
Aspect BISx module (Optional)
Parameter Measurement:
BC:
0~30 (Only limited to the combined use of an external sensor with a BIS module)
EMG:
30~55dB (bar chart) with intensity between 30dB and 80dB (tendency chart)
BIS:
0~100
SQI:
0%~100%
SR:
0%~100%
SEF:
0.5 Hz~30Hz
TP:
40~100 Db
EEG Measurement:
Input impedance >5 MΩ
Noise (RTI)
<2μV (0.25~50 Hz)
Input signal range:
±1 Mv
EEG bandwidth between:
0.25 Hz~110 Hz
NMT Tof-Watch
®
SX (Optional)
Microprocessor-controlled
Stimulation Mode:
TOF, TOFS, PTC, 1Hz Twitch, 0.1Hz Twitch, DBS DBS3.3 and 3.2 (Double Burst), Tetanic Stimulation (Burst), 5s – 50 Hz or 100 Hz
Output
(accuracy ±5% of full scale value)
Surface electrodes:
Constant current,0-60mA(0-12/18μC) up to 5 KOhm.
Monophasic, 200 μs or 300 μs pulse width
Needle electrodes:
Constant current, 0-6 mA (0-0.24 μC) up to 5 KOhm.
Monophasic, 40μs pulse width Acceleration transducer: Accuracy ±5% of full scale value
Temperature sensor:
Range 20.0-41.5 °C (accuracy ±5 °C)
Operation Environment
Power:
AC 100-250 V, 50/60 Hz
Temperature:
5-40 °C
Humidity:
<80%
Patient Range:
Adult, Pediatric, Neonate
Va rugam sa ne contactati pentru oferta de pret!