ATI » CETUS x15 Critical Care Patient Monitor

CETUS x15 Critical Care Patient Monitor

Vizualizari:   1539

Cod: 681
Producator: Axcent Germania
Cod producator: CETUS x15 Critical Care Patient Monitor
In stoc: Nu
Garantie: 24 luni
0.00Lei
Comanda rapida arrows
Nume:
E-mail:
Telefon:
Cantitate:

Detalii produs Features
15.6” High resolution TFT LCD Touch screen
10 waveform display, up to 12-lead ECG analysis
Powerful calculation (Hemodynamic, Drug Dose, Oxygenation, Ventilation)
Pacemaker detection
ST & arrhythmia analysis
SpO2 support PVI and PI, low perfusion 0.2%
Aspect BISx module, NMT (Organon TOF-Watch®) module optional
Wired/Wireless CMS, support HL7 protocol to HIS
SpO2 pulse-tone modulation (Pitch Tone)
VGA support external display
Graphical & tabular trend review (120 hours)
48 hours full disclosure wave review for each patient (stored in SD card)

Configuration
5-lead ECG, SpO2, NIBP, TEMP, Resp, PR; Li-ion battery

Optional
Touch-Screen, 12-lead ECG, Masimo/Nellcor SpO2, IBP, C.O., EtCO2, Multi-Gas, BIS, NMT, VGA, Thermal Recorder, Wired/Wireless CMS


TEHNICAL SPECIFICATIONS

Display

15.6” TFT (touch screen optional)
Resolution: 1366 x 768
Number of traces: 12 waveforms


I/O
LAN: 1 standard RJ45 port
WLAN: IEEE 802.11b/g/n
USB: 2 USB connectors
SD: 1 SD card socket
VGA: 1 VGA monitor connector
Output: 1 connector for Nurse call, Defib Sync Analog Output


ECG
Lead type: 3-lead, 5-lead,12-lead
ECG waveform: 2 channels, 7 channels, 12 channels
Display sensitivity: 2.5 mm/mV (×0.25), 5 mm/mV (×0.5), 10 mm/mV (×1.0), 20 mm/mV (×2.0)
Wave sweep speed: 6.25 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s
Bandwidth
Diagnostic mode: 0.05 Hz~100 Hz
Monitor mode: 0.5 Hz~40 Hz
Surgery mode: 1 Hz~20 Hz
Strong filter mode: 5Hz~20 Hz
CMRR>100 dB

Notch: 50/60 Hz notch filter can be set to on or off
Differential input impedance >5MΩ
Electrode polarization voltage range: ±400mV
Baseline recovery time <3s after defibrillation (in monitor and surgery mode)
Calibration signal: 1 mV (peak - peak), accuracy ±3%


RESP
Measurement method: Thoracic electrical bioimpedance
Measuring lead: Lead I, II
Wave gain: ×0.25, ×0.5, ×1, ×2
Respiratory impedance range: 0.5-5 Ω
Baseline impedance: 500-4000 Ω
Gain: 10 grades
Scan speed: 6.25 mm/s, 12.5 mm/s, 25 mm/s


TEMP
Accuracy: ±0.1 or ±0.2 °F (without probe)
Measurement range: 5~50? (41~122 °F)
Channel: Two channels
Resolution: 0.1?
Parameters:?T1, T2 and TD


SpO2
Measurement range: 0-100%
Resolution: 1%
Accuracy: ±2% (70-100%, Adult/Pediatric); ±3% (70-100%, Neonate); 0-69%, unspecified
Refreshing Rate: 1s


Masimo SET® SpO2(Optional)
Measurement range: 0-100%
Resolution: 1%
Accuracy: ±2% (70-100%, Adult/Pediatric, non-motion, low prefusion); ±3% (70-100%, Neonate, non-motion); ±3% (70-100%, motion); 0-69%, unspecified
Refreshing Rate: 1s


Pulse Rate
Range: 30~300 bpm
Resolution: 1 bpm
Accuracy: ±2bpm (non-motion); ±5bpm (motion)
Refreshing rate: 1s


NIBP
Measurement method: Automatic oscillometric method
Operating mode: Manual, automatic, continuous
Measurement unit: mmHg/kPa selectable
Typical measurement time: 20~40 s
Measurement type: Systolic, Diastolic,
Mean Measurement range (mmHg)

Range of Systolic pressure:
Adult 40-270

Pediatric 40-200
Neonatal 40-135
Range of Diastolic pressure:
Adult 10-210

Pediatric 10-150
Neonatal 10-95

Range of Mean pressure:
Adult 20-230
Pediatric 20-165
Neonatal 20-105
Measurement accuracy
Maximum average error: ±5 mmHg
Maximum standard deviation: 8 mmHg
Resolution: 1 mmHg
Interval: 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180, 240, 480 minutes
Overpressure protection: Software and hardware, double safety protection
Cuff pressure range: 0-280 mmHg


IBP (Optional)
Channel: 2-channel or 4-channel
ART: 0 to 300 mmHg
PA: -6 to 120 mmHg
CVP/RAP/LAP/ICP: -10 to 40 mmHg
Measurement range: P1/P2 -50 to 300 mmHg
Resolution: 1mmHg
Accuracy: ±2% or ±1mmHg, whichever is greater (without sensor)
Sensitivity: 5uV/mmHg/V
Impedance range: 300 to 3000 Ω


C.O. (Optional)
Method: Thermodilution
Range: C.O.: 0.2 to 20 L/min
TB: 23 to 45 ?
T1: -1 to 27?
Accuracy: C.O.: ±5% or ±0.1L/min, whichever is greater TB, T1?±0.5? (without sensor)


Standard Mainstream CO2 (Optional)
Measurement range: 0-19.7%, 150 mmHg, or 0-20 kPa
Resolution: 0.1 mmHg
Measurement accuracy:
0-40 mmHg: ± 2 mmHg
41-70 mmHg: ± 5% of reading
71-100 mmHg: ± 8% of reading
101-150 mmHg: ± 10% of reading
Respiration rate: 3-150 bpm
Respiration rate accuracy: 1 ±1bpm
Warm-up time: 97% within 8 s, full accuracy within 20 s


Standard Sidestream CO2 (Optional)
Measurement rage: 0-20% (0-150 mmHg)
Accuracy:
< 5.0% CO 2: ± 2 mmHg
> 5.0% CO 2: < 6% of reading
Respiration rate: 2~150 BPM
Respiration rate accuracy: 1% ±1BPM
Warm-up time: 97% within 45 s, full accuracy within 10 min.
Rise times (t 10-90%): About 100 ms, when flow is 100 ml/min, adult water trap, 1.5 m sampling tube
Delay time: <3 sec when flow is 100 ml/min, adult water trap, 1.5 m sampling tube


Recorder (Optional)
Built-in, Thermal dot array
Horizontal resolution: 16 dots/mm (25 mm/s paper speed)
Vertical resolution: 8 dots/mm
Paper speed: 25 mm/s, 50 mm/s
Number of waveform channels: 3

Phasein ISA Sidestream CO2 (Optional)
Warm-up time: Full accuracy within 10 seconds
Sampling flow rate: 50ml/min(+/-10/min)
Measurement Range: 0-25%
Accuracy:
0~15% (±0.2% of the reading)
15~25%, unspecified
Rise time: 200 ms, typical at 50 ml/min flow rate
Total response time: within 3 seconds (with 2 m Momoline sampling line)
AWRR Range: 0-150 bpm
AWRR Accuracy: ±1 breath


Phasein IRMA™ Mainstream CO2 (Optional)
Measurement Range: 0-25%
Accuracy:
0~15% (±0.2% of the reading)
15~25%, unspecified
Warm-up time: Full accuracy within ‚ 10 seconds
AWRR Range: 0-150 bpm
AWRR Accuracy: ±1 breath


Phasein IRMA™ AX+ Mainstream Multi-gas (Optional)
Gas: CO2, N2O, HAL, ISO, ENF, SEV, DES with automatic identification
Warm-up time: Full accuracy within 20 seconds for IRMA
AX+ CO2 Accuracy:
0-10%: ± (0.2%+2% of the reading)
0-15%: ± (0.3%+2% of the reading)
N2O Accuracy:
0-100%: ± (2%+2% of the reading)
HAL, ISO, ENF:
0-8%: ± (0.15%+5% of the reading)
SEV:0-10%: ± (0.15%+5% of the reading)
DES:0-22%: ± (0.15%+5% of the reading)
Agent identification time: <20 s (typical <10 s)
AWRR range: 0-150 bpm
AWRR accuracy: +/-1bpm
Apnea time: 20~60 s


Aspect BISx module (Optional)
Parameter Measurement:
BC: 0~30 (Only limited to the combined use of an external sensor with a BIS module)
EMG: 30~55 dB (bar chart) with intensity between 30 dB and 80 dB (tendency chart)
BIS: 0~100
SQI: 0%~100%
SR: 0%~100%
SEF: 0.5 Hz~30 Hz
TP: 40~100 Db
EEG Measurement:
Input impedance >5 MΩ
Noise (RTI) <2 μV (0.25~50 Hz)
Input signal range: ±1 Mv
EEG bandwidth between: 0.25 Hz~110 Hz


NMT Tof-Watch® SX (Optional)
Microprocessor-controlled
Stimulation Mode: TOF, TOFS, PTC,
1 Hz Twitch, 0.1 Hz Twitch, DBS DBS3.3 and
3.2 (Double Burst), Tetanic Stimulation (Burst),
5s – 50 Hz or 100 Hz
Output (accuracy ±5% of full scale value)
Surface electrodes:
Constant current, 0-60 mA (0-12/18μC) up to 5 KOhm.
Monophasic, 200 μs or 300 μs pulse width
Needle electrodes:
Constant current,0-6 mA (0-0.24 μC) up to 5 KOhm.
Monophasic, 40 μs pulse width

Acceleration transducer: Accuracy ±5% of full scale value
Temperature sensor: Range 20.0-41.5°C (accuracy ±5 °C)


Operation Environment
Power: AC 100-250 V, 50/60 Hz
Temperature: 5-40 °C
Humidity: <80%
Patient Range: Adult, Pediatric, Neonate