ATI » CETUS x12 Patient Monitor

CETUS x12 Patient Monitor

Vizualizari:   1754

Cod: 680
Producator: Axcent Germania
Cod producator: CETUS x12 Patient Monitor
In stoc: Nu
Garantie: 24 luni
0.00Lei
Comanda rapida arrows
Nume:
E-mail:
Telefon:
Cantitate:

Detalii produs Features
12.1” color TFT LCD screen
8 waveform display, up to 12-lead ECG analysis
Powerful calculation (Hemodynamic, Drug Dose, Oxygenation, Ventilation)
Pacemaker detection
ST & arrhythmia analysis
OxyCRGs screen
Wired/Wireless CMS, support HL7 protocol to HIS
SpO2 pulse-tone modulation (Pitch Tone)
MEWS (Modified Early Warning Score)
Graphical & tabular trend review (120 hours)
Rechargeable Lithium-Ion Battery (2600 mAh)

Configuration
ECG+SpO2+NIBP+2TEMP+PR+RESP, Li-ion battery

Optional
Touch-Screen, 12-lead ECG, Masimo/Nellcor SpO2, IBP, C.O., EtCO2, Multi-Gas, BIS, NMT, VGA, Thermal Recorder, Wired/Wireless CMS


TEHNICAL SPECIFICATIONS

Display

12.1” TFT (touch screen optional)
Resolution: 800 x 600
Number of traces: 8 waveforms


ECG
Lead type:
3-lead, 5-lead, 12-lead
ECG waveform:
2 channels, 7 channels, 12 channels
Display sensitivity:
2.5 mm/mV (×0.25), 5 mm/mV (×0.5), 10 mm/mV (×1.0), 20 mm/mV (×2.0)
Wave sweep speed: 6.25 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s
Bandwidth:
Diagnostic mode: 0.05 Hz~100 Hz
Monitor mode: 0.5 Hz~40 Hz
Surgery mode: 1Hz~20Hz
Strong filter mode: 5 Hz~20 Hz
CMRR >100 dB
Notch: 50/60 Hz notch filter can be set to on or off
Differential input impedance >5MΩ
Electrode polarization voltage range: ±400 mV
Baseline recovery time <3s after defibrillation (in monitor and surgery mode)
Calibration signal: 1mV (peak - peak),accuracy ±3%


RESP
Measurement method: Thoracic electrical bioimpedance
Measuring lead: Lead I, II
Wave gain: ×0.25, ×0.5, ×1, ×2
Respiratory impedance range: 0.5-5Ω
Baseline impedance: 500-4000Ω
Gain: 10 grades
Scan speed: 6.25 mm/s, 12.5 mm/s, 25 mm/s


Pulse Rate
Range: 30~254 bpm
Resolution: 1bpm
Accuracy: ±2bpm (non-motion), ±5bpm (motion)
Refreshing rate: 1s


TEMP
Accuracy: ±0.1 °C or ±0.2 °C °F (without probe)
Measurement range: 5~50 °C (41~122 °F)
Channel: Two channels
Resolution: 0.1 °C
Parameters: T1,T2 and TD


NIBP
Measurement method: Automatic oscillometric method
Operating mode: Manual, automatic, continuous
Measurement unit: mmHg/kPa selectable
Typical measurement time: 20~40 s
Measurement type: Systolic, Diastolic, Mean
Measurement range (mmHg)
Range of Systolic pressure:
Adult 40-270
Pediatric 40-200
Neonatal 40-135

Range of Diastolic pressure:
Adult 10-210
Pediatric 10-150
Neonatal 10-95

Range of Mean pressure:
Adult 20-230
Pediatric 20-165
Neonatal 20-105
Measurement accuracy
Maximum average error: ±5 mmHg
Maximum standard deviation: 8 mmHg
Resolution: 1 mmHg
Interval: 1, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180,
240, 480 minutes

Overpressure protection: Software and hardware, double safety protection
Cuff pressure range: 0-280 mmHg


SpO2
Measurement range: 0-100%
Resolution: 1%
Accuracy: ±2% (70-100%, Adult/Pediatric); ±3% (70-100%, Neonate); 0-69%, unspecified
Refreshing Rate: 1s


Masimo SET® SpO2 (Optional)
Measurement range: 0-100%
Resolution: 1%
Accuracy: ±2% (70-100%, Adult/Pediatric, non-motion, low prefusion); ±3% (70-100%, Neonate, non-motion); ±3% (70-100%, motion); 0-69%, unspecified
Refreshing Rate: 1s


Recorder (Optional)
Built-in, Thermal dot array
Horizontal resolution: 16 dots/mm (25 mm/s paper speed)
Vertical resolution: 8 dots/mm
Paper speed: 25 mm/s, 50 mm/s
Number of waveform channels: 3


Operation Environment
Power: AC 100-250V, 50/60Hz
Temperature: 5-40 °C
Humidity: <80%
Patient Range: Adult, Pediatric, Neonate


Va rugam sa ne contactati pentru oferta de pret!
Echipa SANTIMPEX