Numar canale EEG |
25 |
Numar canale poligrafice |
4 |
Numar canale respiratie |
4 |
Canale EEG |
|
Esantion de volum/canal |
pana la 5000 Hz |
Sensibilitate |
1 - 500 µV/mm |
Gama intrare |
13mV |
Filtrul trece sus |
0.05 - 2 Hz |
Filtrul trece jos |
15, 30, 75, 100, 200 Hz |
Convertor A/D |
16 bit |
Rejectia in mod normal - frecv. 10Hz |
110 dB |
Rata de atenuare a filtrului notch |
40 dB |
Nivelul zgomotului varf-varf masurat la
rezistenta de 10 kΩ |
0.8 microV |
Impedanta de intrare |
100 MOhm |
Atenuare intre canale, nu mai mica de |
60 dB |
Calibrare |
Auto |
Controlul electrozilor prin masurarea
impedantei |
Da |
Controlul electrodului de
impamantare |
Da |
Canale Poligrafice |
|
Frecventa de esantionare pe canal |
pana la 40000 Hz |
Filtrul trece sus |
0.05-200 Hz |
Filtrul trece jos |
35-5000 Hz |
Sensibilitate |
10-10000 µV/mm |
Gama intrare |
0.2 - 100 mV |
Fotostimulator |
|
Durata stimulului |
20 - 1000 ms |
Frecventa stimulare |
0.1 - 50Hz |
Stimulare Stanga / Dreapta/
Ambele parti |
Da |
Stimulator pattern |
|
Frecventa stimulare |
1 - 10 s |
Dimensiune pattern |
4?3, 8?6, 16?12, 32?24, 64?48 celule |
Fonostimulator |
|
Nivel stimulare |
0 - 110 db |
Ton stimulare |
|
Frecventa |
100 - 2000 Hz |
Durata stimulului |
0.5 - 500 ms |
Stimulare click |
|
Durata stimulului |
100 - 5000 µs |
Frecventa stimulare |
0.1 - 50 Hz |
Stimulare Stanga / Dreapta /
Ambele parti |
Da |
Compresie/Decompresie |
Da |
Canal respiratie |
|
Latime banda |
0.05 - 35 Hz |
Parametrii si caracteristici de baza |
|
Interfata |
USB |
Computer cu configuratia minima |
12,6 GHz, video 512 Mb, HDD 80Gb,OS- Windows XP Professional, monitor 17 inch, tastatura, mouse, CD-ROM -RW |
Imprimanta cu jet de cerneala |
Da |
|
|
|
DESCRIPTION |
|
|
SP700003-25 |
SIGMA PLpro Kit
for 25 channel EEG |
- digital 25 channel amplifier,
21 EEG and 4 polygraphic cannels
- medical power supply
- optical coupling device, fiber optic
cable 5 m with serial port adapter
- complete accessories
(electrodes, cable ...)
- stand for head box
- EEG software package according
to the attached description |
|
|
SP710013 |
SIGMA PLpro EEG
Viewer/Creator |
for free installation on each PC
- creation of EEG database
for file export
- EEG database for selecting
files for review
- EEG review software incl.
meassurement, printout,
EEG comparison, montage selection,
report editor
- CD creation incl. EEG Viewer and
up to 100 EEG´s for external EEG
reviewing or demonstration |
|
|
SP700031 |
Photic stimulator
FSA 10-2L |
additional unit for paperless EEG, 1,0-60 Hz, optical trigger input (optional TTL), LED flash lamp |
|
|
SP700037 |
Stand for Photic
Stimulator FSA |
|
|
|
SP720011 |
Trolley for SIGMA PLpro |
|
|
|
SP720070 |
Isolation transformer 440 VA |
1,9 A, 220 V, 50 Hz or other specific. |
|
|
SP720012 |
Ink Jet printer |
Canon PixmaIP 2000 |
|
|
SP700040 |
SIGMA PLpro PC station |
- Main processor unit min.
INTEL 2,6 GHz, 512 MB RAM,
80 GB HD, CD-RW, keyboard,
mouse, mousepad
- interfaces: network, sound, USB,
parallel, serial, PS2
- OS WINDOWS XP Prof. english
- 17" TFT display
- EEG software installed |
|
|
SP7000 |
Accessory |
21 x Bridge electrodes
21 x EEG connector cables
2 x ECG wrist electrode
2 x ECG connector cables
2 x Ear electrodes
1 x EEG cap |
|